Product Description
- Input : 150~270 VAC , 50Hz
- Battery - 12V, 4.5Ah, Rechagable Battery Capacity 3 hours For monitor & 70~100 Shocks on single Charge
- Charging time less than 6 Seconds
- Operation-by Front Panal Key board, Encoder & ON-Off By front Panel Push Button
- Display : 7" Colour TFT with ECG display and printer / Recorder
- Input : 150 ~ 276 VAC, 50 Hz
- Output : 17V at 4A
- Power Consumption : 30 Watts (Max) in ECG mode
- Operating Temp : -5°C to +50°C
- Humidity : up to 95% non condensing
- Rechargeable lithium battery with charger (11.1V)
- LED Indicator : Low battery and battery charging
Additional General Features
- Operating Mode : Manual & AED
- Paper type Printer : Thermal paper
- SpO , NIBP, Pacemaker, EtCO is optional
ECG
- Lead Selection for ECG : Lead I, II and III
- Sweep speed for ECG : 12.5, 25, 50mm/s
- Waveform 1 : ECG
- Waveform 2 : 4 sec delayed (Cascaded)
- Freeze Mode : To Freeze up to 4 Second ECG
- Frequency response : 0.5Hz to 35Hz
- Patient Isolation : Optical isolation
- Gain Setting : 0.5, 1 and 2 CMRR: ≥90 dB
- Digital Filter : Digital Signal processing for good quality of ECG
- Dimension : 280(L) X 330(W) X 200(H)mm
- Weight : 6.5 Kg
Printer
- Paper Type : Thermal paper50/55mm width,
- 20 meters length for 2 inch printer
- Event Printing : Prints 5 seconds pre & post ECG with Date, Time, Energy and mode, Shock status
- ECG in Defib Mode : Prints last 10 sec ECG through paddle
- Print Annotates in Difib Mode : Time, Date, Mode, Lead name, Gain, Sweep Speed
- ECG in Monitor Mode : Prints last 10 sec ECG through paddle
- Print Annotates in Monitor Mode : Time, Date, Selected Lead name, Gain, Sweep Speed
- Manual Defib Check Mode : Print the status of Shock, Energy, Date and Time
AED Mode
- Waveform : Bi Phasic
- AEC Shock series : Energy level from 150 to 200J
- Charge command automatic after identifying arrhythmia
- Shock Command : Front panel button, Shock
- Charging Time : Less than 7 seconds
- Controls : Analyze, Pause
- Defibrillation Electrode : Malfunction, Self Adhesive, Adult electrodes
- Voice and Text Prompts : Voice and text prompts to guide user
Pacemaker (Optional)
- Waveform: Monophasic
- Modes: Demand or Non Demand
- Amplitude: From 10mA to 150mA (Resolution of 10mA)
- Pulse Width: 40ms (± 1ms)
- Frequency: From 20ppm to 250ppm (increments of 1ppm)
- Refectory period: 340ms (from 30 to 80ppm), 240ms (from 90 to 180ppm)
- Maximum Output Voltage: 150V
- Impedance Range: 200Ω to 1kΩ
SpO2 (Optional)
- Range: 0 ~ 100%
- Accuracy: 70 ~ 100% ±2%, 50 ~ 69% ±3%
- Pulse rate: Displayed when PR is deselected
- Pulse Beep: Step less volume control
- PR Range: 10 ~ 250bpm
- Pitch Variation with change in oxygen saturation
NIBP (Optional)
- Method : Oscillometric
- NIBP Interval : User Selectable - 3 min to 120 min
- Modes : Manual, Auto
- Cuff Deflation : Automatic
- Safety Features : Automatic deflation if cuff pressure exceeds 280mmHg
- Range : 10 ~ 250mmHg
- NIBP Status : “Ready” - When not taking reading
- “Measuring” - When taking reading
- “Error” - If reading could not be taken due to some error while measuring
- Patient Type : Adult, Paediatric, Neonate