- 8″ LCD display with Touch screen
- Power Supply:
- AC 100 ~ 240V, 50/60Hz
- DC 14.8V, 3700mAh rechargeable lithium battery
- Input way: Floating and Defibrillation Protection
- Lead: Standard 12 leads
- Patient leak current : ≤ 10μA
- Input Impedance : ≥50MΩ
- Frequency Response : 0.05 ~ 150Hz [-3dB ~ 0.4dB]
- Time Constant : Time ≥3.2s
- CMRR : ≥ 60dB, ≥100dB [Add Filter]
- EMG Interference Filter : 35Hz [-3dB]
- Recording way : Thermal Printing System
- Recording: Auto, Manual and Rhythm
- Paper Size: 210mm [W] x 20m [L] high speed thermal paper
- Paper Speed:
- Auto Record: 25mm/s, 50mm/s, error ±5% [Record setup according to auto record format and mode, automatically changing leads, measuring and analyzing]
- Rhythm Record: 25mm/s, 50mm/s, error ±5% [Record setup according to auto record format and mode, automatically measuring and analyzing]
- Manual Record: 5mm/s, 6.25mm/s, 10mm/s, 12.5mm/s, 25mm/s, 50mm/s, error ±5% [Record setup according to auto record format and mode, manual changing leads]
- Sensitivity Selection : 5, 10, 20, 40 mm/mV, [Error ±5%]
- Standard sensitivity is 10mm/mV ±0.2mm/mV
- Printing Modes:
- 12*1,
- 6*2 + 1,
- 6*2,
- Rhythm row 12, row 10, row 8, row 6 etc with multiple formats,
- Measurement parameters : HR, P-R interval P duration, QRS duration, T duration, Q-T interval, Q-Tc, P Axis, QRS Axis, T Axis, R[V5], S[V1], R[V5]+S[V1]
- Product Safety Type: Class I CF applied part, there is defibrillation and pacing protection circuit in it
- Ending Polarization voltage: ±5OOmV
- Noise Level: ≤15μVp-p
- Machine Size: 334mm[L] X 320mm[W] X 85mm[H]
- Machine Net Weight: 2.25Kg
OPTIONAL: Communicate with PC via RS 232 interface and Software>